ACQUITY pump unable to compress - WKB108550
Article number: 108550
SYMPTOMS
- Pump can prime—solvent comes out of the waste line
- No visible leaks at pump tubing or check valves
- When dynamic leak test starts, accumulator pump is unable to compress
- Replacing the high-pressure seals does not resolve the issue
- Low pressure (3 psi) is seen in the console when setting flow to 0.5 mL/min through the column
ENVIRONMENT
- ACQUITY UPLC I-Class BSM
- ACQUITY UPLC BSM
- ACQUITY UPLC H-Class QSM
CAUSE
Air in the pump head is preventing solvent from moving through the pump.
FIX or WORKAROUND
- Remove the stainless steel fitting under the transducer on the primary pump.
- Set the flow to 0.5 mL/min. Eventually, solvent will flow through the bottom of the transducer:
- Rerun the leak test.
ADDITIONAL INFORMATION