With ACQUITY Arc, baseline ripples during gradient - WKB108324
Article number: 108324
SYMPTOMS
- The baseline ripples during the gradient
- Leak test successful
- No pressure abnormalities seen during run
ENVIRONMENT
- ACQUITY Arc
- 2998 Photodiode Array Detector
- 2489 UV/Vis Detector
CAUSE
The pump stroke volume is high.
FIX or WORKAROUND
- When using the Path 1 flow path, set the stroke volume to 132 μL (default).
- When using the Path 2 flow path, set the stroke volume to 66 μL.
ADDITIONAL INFORMATION
