High area %RSD using SMFTN - WKB107852
Article number: 107852
SYMPTOMS
- High area %RSD observed
- Initial injections yield consistent peak areas
- Peak area becomes erratic after a few injections
ENVIRONMENT
- ACQUITY UPLC H-Class
- ACQUITY UPLC H-Class Bio
- Acquity Arc
- Arc HPLC
- Acquity UPLC I-Class with SMFTN
CAUSE
Insufficient Sample Manager purge during system setup
FIX or WORKAROUND
- Open the instrument console.
- Navigate to Sample Manager FTN in the system tree.
- Select Control from the drop-down menus along the top of the window.
- Select Prime from the drop-down menu.
- Ensure that the Purge Solvent selection is made.
- Specify the appropriate number of cycles to be performed. (At least 40 cycles are recommended for a complete changeover of solvents.)
ADDITIONAL INFORMATION
id107852, A-30SM, UPISMFTN, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO