Skip to main content
Waters

2545 Binary Gradient Module fails the static decay test - WKB102177

Article number: 102177

SYMPTOMS

  • One head fails the static decay test at 5000 psi
  • Dramatic failure in pressure when running the leak test
  • Test fails regardless of solvent
  • New check valves, seals, head, wash housing, and plunger make no difference

ENVIRONMENT

  • 2545 Binary Gradient Module

CAUSE

Leak on the transducer outlet tubing to the solvent selector manifold

FIX or WORKAROUND

Replace the tubing.

ADDITIONAL INFORMATION

Left head tubing: part 430001300

Right head tubing: part 430001298

Not able to find a solution? Click here to request help.