Ghost peaks in isocratic methods with 2695 - WKB100863
Article number: 100863
SYMPTOMS
- Ghost peaks occur after sample injection
- Ghost peaks occur even after sufficient needle wash
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
CAUSE
Sample stuck to the needle and difficult to remove
Sample carryover
FIX or WORKAROUND
- Fill the vial with a solvent that dissolves the sample well and inject it. (50ul x 10 times)
- If the peak becomes smaller by repeated injections, repeat step 1.
ADDITIONAL INFORMATION
