Ghost peaks in isocratic methods with 2695 - WKB100863

Article number: 100863

SYMPTOMS

  • Ghost peaks occur after sample injection
  • Ghost peaks occur even after sufficient needle wash

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module

CAUSE

Sample stuck to the needle and difficult to remove

Sample carryover

FIX or WORKAROUND

  1. Fill the vial with a solvent that dissolves the sample well and inject it. (50ul x 10 times)
  2. If the peak becomes smaller by repeated injections, repeat step 1.

ADDITIONAL INFORMATION