ACQUITY UPLC degassser vacuum hw error - WKB10065

Article number: 10065

SYMPTOMS

  • Degassser vacuum hw error after power-on.
  • Degasser no vacuum error
  • Reset does not help.
  • Restart does not help.

ENVIRONMENT

  • ACQUITY UPLC BSM
  • ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
  • ACQUITY UPLC I-Class

CAUSE

  • System not in use for a period of time.
  • Possible air in system.

FIX or WORKAROUND

  1. Prime pump all lines with 100% methanol for 15-20 minutes.
  2. Restart pump 
  3. Reset pump in ACQUITY console.

ADDITIONAL INFORMATION