ACQUITY Arc: Area value is smaller than before when returning from a long idle time - WKB100534
Article number: 100534
SYMPTOMS
- The area values decrease to about 1/4 compared to before when returning from a long idle time
- Visual check of bubbles in the sample syringe
ENVIRONMENT
- ACQUITY Arc
- ACQUITY Arc Sample Manager FTN-R
CAUSE
Insufficient purging of the sample manager’s sample syringe. The specified amount cannot be injected due to the air bubbles.
FIX or WORKAROUND
- Purge the sample syringe until air bubbles are removed
- If the bubbles persists, first purge with methanol until the bubbles are removed
- Purge with sufficient solvent used for measurement