Skip to main content
Waters

2707 not found in Breeze 2 node properties - WKB10001

Article number: 10001

SYMPTOMS

  • 2707 cannot be found in Breeze 2 node properties

ENVIRONMENT

  • 2707 Autosampler
  • Breeze 2

CAUSE

Problem with 2707 ICS 

FIX or WORKAROUND

  1. Ensure that all Ethernet cables are connected correctly
  2. Determine the 2707 ICS version by clicking Control Panel > Programs > View Programs.
  3. Uninstall 2707 ICS and restart the computer.
  4. Install 2707 ICS and restart the computer.
  5. Log in to Breeze 2 and connect again.

ADDITIONAL INFORMATION

2707 Autosampler Instrument Control Software Version 1.20

id10001, 2707, 2707SFC, BREEZESW, SUPBZ

Not able to find a solution? Click here to request help.