How to resolve communication issues with Waters Tof MS instruments (Vion, Xevo, RDa) in a waters_connect network - WKB279461
Article number: 279461
OBJECTIVE or GOAL
Resolve communication issues with Waters Tof MS instruments (Vion, Xevo, RDa) in a waters_connect network.
ENVIRONMENT
- waters_connect versions since 3.2.0 LTS
- Network
PROCEDURE
- Ensure that no other users have the System Console App open.
- Reset the MS from the Instrument pane in the System Console App. Move to the next step if this does not work.
- Close the System Console App.
- Open the Device Management App and reset the LND.
- If it reports "LND not available" or nothing happens, continue to step 7.
- If the LND resets correctly and comes back online, move to step 11.
- Connect via remote desktop to LND (if permitted) and stop / start the waters_connect LND service (This has been reported for Vion to be required after Resets fail)
- If the LND resets correctly and comes back online, move to step11.
- Physically reboot the LND computer next to the system.
- In Device Management, confirm that the LND comes online successfully.
- Do a soft reset of the MS. This is usually done with a small recessed button on the front of the instrument.
- Determine if the MS is now communicating in the System Console.
- If not, try a full power reset of the MS using pump override.
- Determine if the MS is now communicating in the System Console.
- You may also need to repeat the LND reset if there are still comms issues.
- Finally, the MS, depending on the instrument will need settting up again and recalibrating.
- If this fails repeatedly, raise a request for Waters Support. We will need to look for an issue with the MS EPC startup or IT security settings in the customer environment.
ADDITIONAL INFORMATION