Deletion of a parent folder also deletes export/delete events of the child folders that were previously removed (CRI-4828) - WKB242935

Article number: 242935

SYMPTOMS

  • Deletion of a parent folder also deletes the export/delete events of the child folders that were previously removed

ENVIRONMENT

  • UNIFI 1.9.4 and earlier
  • waters_connect/UNIFI 

CAUSE

  • CRI-4828
  • Reproduced in waters_connect UNIFI 3.0.0

FIX or WORKAROUND

In order to work around this issue, Waters advises that the data folders be structured so that export/delete for data archival is done at the parent level instead of the child folder level.