MassLynx imports fail during import stage in UNIFI app in waters_connect UNIFI 1.9.13 - WKB237259
Article number: 237259
SYMPTOMS
- Error: Capture operation was canceled due to conflicts with another program concurrently making changes to the data
- Error: Error writing stream: Object reference not set to an instance of an object
- Error: Error writing stream: Stream not open
ENVIRONMENT
- waters_connect
- UNIFI 3.1.0 and earlier
- Workstation (Network not affected)
CAUSE
CRI-4694 (INFPLT-40416)
A new data file is not created during a MassLynx import. This causes the various writing stream errors in the database.
FIX or WORKAROUND
- This issue was fixed in UNIFI 3.4.0, which is included in waters_connect 3.1.1.
- Please upgrade
ADDITIONAL INFORMATION
id237259, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC