UNIFI Export to PDF does not work - WKB92044

Article number: 92044

SYMPTOMS

  • When selected Export to PDF then UNIFI just loading for minutes and nothing happens.
    clipboard_e3274fe19085f9f06c34c09b0589bf5cf.png
  • No errors seen.

ENVIRONMENT

  • UNIFI 1.9.4

CAUSE

The drive that is selected to export the PDF to is not available, even though it can be selected.

FIX or WORKAROUND

  1. Try to export to another location e.g. Desktop or local disk.
  2. If 1 works, then try to restart PC and export to another drive.
  3. If works locally and not on external drive, then investigate permissions for the selected drive.

ADDITIONAL INFORMATION

Export to PDF is selected here:
clipboard_eecc039eac76cc0b1a21e7a7b1bb9c766.png