Skip to main content
Waters

How to change extended loop or sample syringe size on ACQUITY SM-FTN in UNIFI / Waters_Connect - WKB81557

Article number: 81557

OBJECTIVE or GOAL

Change extended loop size or sample syringe size.

ENVIRONMENT

  • UNIFI
  • Waters_Connect
  • ACQUITY UPLC I-Class
  • ACQUITY UPLC H-Class

PROCEDURE

  1.  Open the active instrument system from UNIFI.
    1. Open System Console if using Waters_Connect
  2.  Navigate to the Sample Manager tab, select Setup, and then select Configure.

Configure extension loop.png

  1.  Click on the edit icon (on the far-right of the window) to open the Edit Settings dialog box.
    clipboard_e0dce06272387af88dd7d99c83a353058.png
  2.  Input new settings for Extended loop size or Sample syringe size.

Configure extension loop 2.png

  1.  Select OK.

ADDITIONAL INFORMATION

 

id81557, A-30SM, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN, UPISMFTN+, UPPFTN, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN, UPSMFTN+, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO

Not able to find a solution? Click here to request help.