Random instrument EPC failure or loss-of-communication error in UNIFI with Xevo G2-S - WKB66064
Article number: 66064
SYMPTOMS
- Loss of communications
- Rebooting the EPC restores communications
ENVIRONMENT
- UNIFI
- Xevo G2-S
CAUSE
Using a startup/shutdown method from a previous version of UNIFI
FIX or WORKAROUND
Create a new startup/shutdown method in the current version of UNIFI.