Communication issue with ACQUITY PDA in UNIFI workgroup - WKB51168

Article number: 51168

SYMPTOMS

  • When resetting the PDA, it initially shows green/idle in the console pane, but it then switches to red and shows communication error
  • All other components have good communications

ENVIRONMENT

  • ACQUITY PDA
  • UNIFI Workgroup

CAUSE

Unknown

FIX or WORKAROUND

  1. Take the system offline and close UNIFI.
  2. On the LND, stop the Waters Service Host service and restart it; ensure that it restarts with no errors.
  3. On the UNIFI client, open UNIFI software, navigate to Device Management > Laboratory Network Devices and reset the device.
  4. Bring the system back online.
  5. Verify that communications to the PDA are satisfactory.

ADDITIONAL INFORMATION

 