Skip to main content
Waters

How to Manually Pick Peaks in UNIFI's Intellistart Calibration Routine - WKB4526

Article number: 4526

OBJECTIVE or GOAL

Manually match peaks against the expected calibration masses display in UNIFI.

ENVIRONMENT

  • UNIFI 1.8.1

PROCEDURE

  1. After running the Calibration routine, go to the Results pane.
  2. Select the result that shows a status of Ready for Peak Matching.
  3. Right-click the entry and select Peak Match to open the Calibration Peak Matching window.
  4. Click and drag across the area of the spectrum requiring peak matching.
  5. Select a reference peak to be matched so it is highlighted.
  6. Select the appropriate peak in the acquired spectrum; now both the reference and acquired peaks are highlighted.
  7. Continue until all or most peaks are matched.
  8. Click on Create Calibration.
  9. If the residuals are acceptable, click Accept.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.