UNIFI error message: Cannot acquire analysis, the mass range does not agree - WKB3200
Article number: 3200
SYMPTOMS
- When the sample set starts, there is a UNIFI error message: Cannot acquire analysis, the mass range does not agree
ENVIRONMENT
- UNIFI 1.8.1
CAUSE
The calibration mass range and the analysis mass range do not match.
FIX or WORKAROUND
- In the instrument system control, under the MS, go to Setup and configure.
- Be sure to select a calibration that fits the mass range of the analysis being run.
- When a calibration with the correct mass range is selected, click Apply.
- Start.
ADDITIONAL INFORMATION