Error message "Too many intensity drop outs" during detector setup in UNIFI - WKB31207
Article number: 31207
SYMPTOMS
- Pos and/or Neg fails detector setup
- Error message "Too many intensity drop outs"
ENVIRONMENT
- UNIFI
- Xevo G2-S
- Xevo G2-XS
CAUSE
Incorrect settings for the DRE lenses and collision cell.
FIX or WORKAROUND
Use the default settings for the Collision cell and DRE lenses.
Xevo G2-S/XS
Collector = 60
Collector Pulse = 10.0
Stopper = 10
Stopper Pulse = 20.0
Xevo G2-S only
Entrance = 2
Static offset = 120
Exit = 15 in Sensitivity mode, 5 in Resolution mode
T-Wave trap height = 4
T-Wave extract height = 15 in Sensitivity mode, 5 in Resolution mode
Xevo G2-XS only
Entrance DC = 2
Cell Gradient = 1.0
Static Offset DC (Cell Offset A) = 120
Cell Offset B = 0
Cell Offset C = +0.5
Exit = 15 in Sensitivity mode, 5 in Resolution mode
ADDITIONAL INFORMATION