Error "This license type is not a valid software type for your installation" when trying to activate a small-molecule license in UNIFI - WKB25181
Article number: 25181
SYMPTOMS
- Error "This license type is not a valid software type for your installation" when trying to activate a small-molecule license in UNIFI
- Base license is activated
- Base license and small-molecule license are associated with the same customer number
- Base license and small-molecule license are both Workstation
ENVIRONMENT
- UNIFI Workstation
CAUSE
The serial number of the small-molecule license is not unique.
FIX or WORKAROUND
Contact your local Waters support team to have the issue corrected. When the duplication is corrected, the license should activate as intended.
ADDITIONAL INFORMATION