New waters_connect user cannot log in after resetting password - WKB247799

Article number: 247799

SYMPTOMS

  • New user created
  • Password reset before user logs in
  • Generated password does not work

ENVIRONMENT

  • waters_connect 3.0

CAUSE

The default password is still active.

FIX or WORKAROUND

Log in with the username as the password, and then reset the password. 

ADDITIONAL INFORMATION

The password cannot be reset until after a user has logged in at least once. The reset appears to work but the password is not changed.