Frequent buffered data error messages in UNIFI - WKB24142
Article number: 24142
SYMPTOMS
- Every week or two, there is a buffered data error message
- Clearing the buffered data works, but the buffered data error message appears again a week or two later
ENVIRONMENT
- UNIFI
- UNIFI Network
CAUSE
Possibly a corrupt instrument system GUID
FIX or WORKAROUND
- Create a new instrument system with the same components, and run with the new instrument system.
ADDITIONAL INFORMATION
id24142, SUPUNIFI, UNIFISVR, UNIFISW18, UNIFOPT, UNIFQLIC, UNIFSW17, UNIFSW18, UNIFSW19, UNIFWGLIC, UNIFWKLIC