UNIFI 1.9.2 or 1.9.3 Network: Method upgrade or import error - WKB23273
Article number: 23273
SYMPTOMS
- An error occurs when trying to upgrade a method created in a previous version of UNIFI.
- Vion installation method UEP files imported into UNIFI Network system appear corrupted when you open them. No error is reported during UEP import. The same UEP files do not appear corrupted when imported into a UNIFI Workstation system
ENVIRONMENT
- UNIFI 1.9.2
- UNIFI 1.9.3
CAUSE
Defect in the UNIFI network CR is US1#216993.
FIX or WORKAROUND
Upgrade the method on the UNIFI server before opening on the client. Use the "Legacy Methods" document attached as a guide.
ADDITIONAL INFORMATION