UNIFI 1.8 installer prerequisite error: Reboot is required prior to installation - WKB2081

Article number: 2081

SYMPTOMS

 Installer prerequisite error when trying to install UNIFI 1.8 Workstation: "Reboot is required prior to installation"

ENVIRONMENT

  • Windows 7 Professional SP1
  • UNIFI 1.8 Workstation

CAUSE

A Windows registry flag is set that indicates the computer needs to be rebooted.

FIX or WORKAROUND

  1. Run Regedit from the Windows Start menu.
  2. Search from the top of the registry down for instances of the key PendingFileRenameOperations and delete them.
  3. Without rebooting, install UNIFI.

ADDITIONAL INFORMATION

 