Sample set stopped shortly after submission in UNIFI is stuck in acquire - WKB13987
Article number: 13987
SYMPTOMS
- Instrument and analysis are stuck in acquire
- Cannot submit a new sample set because it says "waiting for the previous sample set to complete"
ENVIRONMENT
- UNIFI Workgroup
CAUSE
Sample set was stopped too quickly.
You must wait until the injection has completed and the acquisition has started in order to stop a run.
FIX or WORKAROUND
- Exit the client.
- Power-off the LND.
- Reboot the server.
- Wait for the watservicehosts to start on the server.
- Power-on the LND.
- Log in to the client.