Arc QSM-R stalls during prime and freezes with no error message - WKB291751

Article number: 291751

SYMPTOMS

  • Randomly, about once a week, Arc QSM-R stalls during prime and freezes with no error message
  • You can hear the sound of vent valve rotation, but there is no sound from plunger
  • No red LED 
  • Restarting the pump does not help
  • Restarting the LAC/E is the only solution

ENVIRONMENT

  • DP2022R1
  • Empower 3.7.0 Enterprise
  • Arc HPLC

CAUSE

Corrupted driver

FIX or WORKAROUND

Repair the driver using the Driver Pack Deployment Manager.

ADDITIONAL INFORMATION

 