The Pretreatment Method, Auxiliary Channels, and General tabs are missing in the Instrument Method for the Agilent AGGC using Agilent ICF Support 3.0 - WKB27623
Article number: 27623
SYMPTOMS
Pretreatment Method, Auxiliary Channels, and General tabs are blank.
ENVIRONMENT
- Agilent 7890 GC
- Empower 3
- Agilent Instrument Control Framework (ICF) Support Version 3.0
CAUSE
- PCS 55755 - The Pretreatment tab in the Instrument Method Editor is for LC instrument methods. It is not used in GC instrument methods.
- PCS 55502, 55503: The General tab in the instrument method editor contains information that pertains to LC methods only. When working with instrument methods for GC instruments, the General tab is left blank.
- If there are no Auxiliary options installed on the GC, then there is nothing in this tab.
FIX or WORKAROUND
N/A Expected behavior.