NuGenesis SDMS File Capture does not archive files from a scan path configured using the Ignore Archive Bit Flag - WKB92290
Article number: 92290
SYMPTOMS
- NuGenesis SDMS File Capture is not archiving any files from a particular scan path configured using the Ignore Archive Bit Flag
- Files that are in fact archive candidates are not being identified as such
- The Archive Agent log file records that the scan path is being scanned
ENVIRONMENT
- NuGenesis SDMS 8
- NuGenesis SDMS 9
CAUSE
In some cases, this may be due to a corrupted EVS Dates file for a particular scan path.
FIX or WORKAROUND
- If this issue is due to a corrupted EVS Dates file, stop the Archive Agent, and then delete the corresponding EVS Dates file.This file is located in the Drive:\Program Files (x86)\NuGenesis 8.0\EVS\NugEVSDates folder as a file that corresponds to the path you're looking for.
- Restart the Archive Agent.
ADDITIONAL INFORMATION
Please check with Support/GSS first, before share this KCS with customer
If you use the Ignore Archive Bit and Delete this file, all data will be re-archived again