Error appears when trying to restore NuGenesis SDMS file data in VISION, where data resides on Offline Storage Manager Device - WKB70412
Article number: 70412
SYMPTOMS
- Error received when trying to restore NuGenesis SDMS file data in VISION, where data resides on Offline Storage Manager Device
- Example of error: "The device \\UNCpath\ containing the file..."
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
The NuGenesis RPC service on the NuGenesis Web Server cannot access the data on the OSM device.
FIX or WORKAROUND
- Log in to NuGenesis SDMS Web Server with the RPC Service account and verify that the OSM location is accessible using Windows Explorer.
- If data is visible in the OSM Device share location, verify that the NuGenesis RPC Windows Service is running using the NuGenesis RPC Service user.
- If it is running as "local system" or any other user, stop the service, configure to run as the NuGenesis RPC user, and restart the service.
- Restore the data from NuGenesis VISION again.
ADDITIONAL INFORMATION
See also WKB21005 for an issue that is specific to NG8 SDMS SR1.