The Empower project list displayed in the NuGenesis SDMS template builder does not include all projects in the system - WKB65306
Article number: 65306
SYMPTOMS
- When opening the project exceptions list in SDMS Archive Template Builder, some projects are not shown, even though the Empower user is a member of the Administrators group in Empower
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- Empower 3
CAUSE
The project access rights do not allow access to the named user account.
FIX or WORKAROUND
- If the project's access type is Owner Only, the user account used for SDMS project backup must be the project owner.
- If the project's access type is Owner and Group, the user account must be a member of at least one of the groups that has access to the project.
- If the project's access type is Owner, Group, and World, the World User Type must be non-null.