Error "Access denied" appears when using the NuGenesis LMS server configuration tool - WKB64212

Article number: 64212

SYMPTOMS

  • "Access denied" message appears when saving changes to the LMS server through the configuration tool

ENVIRONMENT

  • NuGenesis 9 LMS
  • NuGenesis 8 LMS

CAUSE

Insufficient file system permissions on the NuGenesis LMS Server installation directory.

FIX or WORKAROUND

Grant the logged-in user Read and Write permissions on the Drive:\WatersLMSServer folder and all subfolders/files.

ADDITIONAL INFORMATION

 