Duplicate document IDs created in NuGenesis LMS - WKB59653
Article number: 59653
SYMPTOMS
- Users are unable to create documents in NuGenesis LMS. The following message appears in LMS when creating a new document:
- Please correct the highlighted errors first!
- When clicking OK to the message box, the document ID field is highlighted in red, and the message text appear when moving the mouse over the field:
- A document with the provided Document ID already exists
- In other cases, the document may be created, but the document ID may be indistinguishable from IDs created by other users
- IDs for reagents or instruments may not be unique if they are based on the document IDs
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- NuGenesis 8 ELN
CAUSE
The document ID pattern does not incorporate enough uniqueness to prevent the generation of duplicate or similar IDs.
FIX or WORKAROUND
- Waters recommends using at least the following elements in the document ID pattern:
- The full date stamp, as in YYYYMMDD
- 3 or more digits for the system-wide document counter or the per-user document counter
- At least 8 characters of the department name, team name, or the user ID
- No Reset on the system and user document counters
ADDITIONAL INFORMATION
Using the full date/time stamp in the ID - YYYYMMDDHHMISS - is often enough to create unique IDs, but tends to create hard-to-remember IDs.
If the department, team, or user ID is incorporated into the ID pattern, then use enough characters to ensure uniqueness across similar names.