The NuGenesis Database Insertion (DBI) plugin does not create trigger files when records are added to the database - WKB51998
Article number: 51998
SYMPTOMS
- The NuGenesis Database Insertion Plugin is installed and configured but it is not creating trigger files
ENVIRONMENT
- NuGenesis 8 SDMS
CAUSE
The configuration program was run without admin rights and the data was not written to the registry; or, the user was printing to a different server than the one named in the DBI Plugin's configuration.
For NuGenesis 8 SR1 only: a hotfix is required for the DBI Plugin to create the trigger files.
FIX or WORKAROUND
- Right-click the shortcut for the DBI Plugin and click "Run as Administrator".
- For NuGenesis 8 SDMS SR1: Contact Waters Technical Support to get a copy of the hotfix.
- Take note of the server name in the DBI Plugin's configuration.
- If the server name that appears in the UNIFY login box does not match the server name in the UNIFY client, use the NG80LocM tool to update the UNIFY client's configuration.