The error "Unable to save template" appears when saving a NuGenesis 8 file capture template - WKB3551
Article number: 3551
SYMPTOMS
- "Unable to save template" appears when trying to save a file capture template
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
CAUSE
The NuGenesis file capture template exceeds the maximum number of lines and modifications.
Note: This issue occurs when you reach approximately 150 template lines and approximately 150 modifications.
FIX or WORKAROUND
- Install the Extract Last Template utility on a machine that has SDMS Administrator installed. This utility is available on the NuGenesis Utilities discs.
- Run the Extract Last Template utility.
- Log in to the utility with an SDMS account.
- Select the project and template, and then click Extract.
- A message box appears stating that the template is saved. You can modify the extracted template in SDMS Administrator.
ADDITIONAL INFORMATION
Contact Waters Technical Support to receive a copy of the Extract Last Template utility.