Error "No Action Available" when editing an MS Office document section in NuGenesis LMS and Acrobat is installed - WKB274017
Article number: 274017
SYMPTOMS
- The following error messages appear when opening an Excel, Word, or PowerPoint document section for viewing or editing in NuGenesis LMS:
- "No Action Available"
- "The selected version cannot be opened. Probably it is an un-editable picture or the default editor is not available"
- The Microsoft Office 365 suite is installed on the LMS client machine
- Adobe Acrobat is installed on the LMS client machine
- No Office sections can be opened
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
CAUSE
The Adobe Acrobat COM Add-in is enabled and interfering with the LMS SmartBuilder Com Add-in.
FIX or WORKAROUND
Disable the Acrobat COM Add-in in Excel on all LMS clients:
-
- File > Options > Trust Center > Add-ins > COM Add-ins from Manage drop-down list > click Go: Clear the selection of Acrobat COM Add-in.