How to enable logging for the Paradigm indexing server - WKB22251
Article number: 22251
OBJECTIVE or GOAL
Configure a Paradigm index server to write the output of an indexing task to a log file.
ENVIRONMENT
- Paradigm Scientific Search 4.2
PROCEDURE
- Copy the information from the Additional Information section into a new text file in Notepad.
- Save the file with the .reg extension.
- Copy the file to the Paradigm index server.
- Double-click the .reg file to add its entries to the registry.
- Restart the Index server.
- Start an indexing pipeline. Paradigm will copy the task's output to the specified log file.
ADDITIONAL INFORMATION
Windows Registry Editor Version 5.00
[HKEY_LOCAL_MACHINE\SOFTWARE\Wow6432Node\Waters\ParadigmScientificSearch]
"TasksPipelineViewFile"="c:\\ParadigmLog\\tasks.log"
"TasksPipelineViewFileSizeLimit"=dword:00004000