Images do not transfer from the Waters Image Writer to NuGenesis LMS - WKB20580
Article number: 20580
SYMPTOMS
- When printing the contents of an LMS document section to the Waters Image Writer, the images do not appear in LMS
ENVIRONMENT
- NuGenesis 9 LMS
- NuGenesis 8 LMS
- NuGenesis 8 ELN
- Sections are routinely opened for external editing. If sections are opened for internal editing (that is, editing within the LMS window), then see "Images are not created when a NuGenesis LMS Smart Builder section is opened for Internal editing"
CAUSE
The wiwagent.exe process is not running.
FIX or WORKAROUND
- For NuGenesis 9 LMS and 8 LMS FR1 and SR2:
- Open the Services control panel (Start > Run > services.msc) and look for the VPDAgent service.
- This service must be started and logged in as LocalSystem.
- If the service is not present, follow the instructions in the article "How to reinstall the printer "Waters Image Writer" for NuGenesis LMS" to install the printer and service.
- For NuGenesis 8 ELN SR1:
- Browse to the ELN client's installation path, typically: Drive:\Program Files (x86)\Waters\NuGenesis ELN\Waters Image Writer
- Double-click the file wiwagent.exe.
ADDITIONAL INFORMATION
For additional considerations when the LMS client runs on a Citrix server, please see the article "Printed images not transferred from the Waters Image Writer to NuGenesis LMS in Citrix sessions".