Print jobs from the NuGenesis Report splitter service are not printed, and jobs remain in the UNIFY printer queues - WKB125295
Article number: 125295
SYMPTOMS
- Print jobs started from the Report Splitter service are stuck in the NuGenesis UNIFY queues with the status "Printed"
ENVIRONMENT
- NuGenesis 9 SDMS
- NuGenesis 8 SDMS
- The NuGenesis SDMS Report Splitter is installed as a service, configured, running, and printing the split reports to the appropriate UNIFY queues
CAUSE
No ngprox.exe or ng80prox.exe process is running for the Report Splitter service account.
FIX or WORKAROUND
- Log in to the Report splitter server with RDP.
- Use the RunAs.exe tool in the command prompt to start the ngprox process:
- NuGenesis 9: runas.exe /user:domain\username "Drive:\Program Files (x86)\Waters\NGSDMS\UV\ngprox.exe"
- NuGenesis 8: runas.exe /user:domain\username "Drive:\Program Files (x86)\NuGenesis 8.0\UV\ng80prox.exe"
- Lock and close the RDP session, but do NOT log off (doing so closes the ngprox process).
ADDITIONAL INFORMATION
Defect report CRI-3146 was filed for this issue.