The NuGenesis SDMS Record Router service is not processing records - WKB101251
Article number: 101251
SYMPTOMS
- No records processed by NuGenesis SDMS Record Router
ENVIRONMENT
- NuGenesis 8 SDMS
- SDMS Record Router v3.x
CAUSE
The Record Router Windows Service is manual and was not started.
FIX or WORKAROUND
- Ensure that the Record Router Windows Service is set to Automatic and that the Service has been started.
- Note that if the service's startup type in Windows is manual, and a user clicks Start Service in the console, the service will process one batch and then stop. In order for the service to continually process batches of records, the service's startup type must be "Automatic" or "Automatic (Delayed start)".