Error "ExecuteTemplateCommand. (5): Argument 'Index' is not a valid value..." appears in the NuGenesis Archive Agent log file - WKB100248

Article number: 100248

SYMPTOMS

  • The following message appears in the NuGenesis SDMS Archive Agent log file:
    • ExecuteTemplateCommand. (5): Argument 'Index' is not a valid value...
  • This error message occurs for each execution of a template line

ENVIRONMENT

  • NuGenesis 9 SDMS
  • NuGenesis 8 SDMS

CAUSE

  1. The scan path in the template line is not a well-formed UNC path.
  2.  Project tags were not set properly as per template requirement.

FIX or WORKAROUND

Solution 1:

  1. Edit the archive template in SDMS Administrator and ensure that the scan path in each template line matches the following format:
    • \\hostname\sharename\optionalFolderName
  2. Save and close the template.
  3. Restart the Archive Agent.

Solution 2:

  1. Ensure that Project tags are correct per the template requirement.
  2. Ensure that all required Template fields are filled. 
  3. Save and close the template. 
  4. Restart the Archive Agent.

ADDITIONAL INFORMATION

 

id100248,

Not able to find a solution? Click here to request help.