FractionLynx collects every fraction even though the threshold is high - WKB91085

Article number: 91085

SYMPTOMS

  • 2767 collects every fraction even though the threshold is high

ENVIRONMENT

  • FractionLynx
  • 2767
  • 2489

CAUSE

Wavelength A or B or C is set to a value other than 0 in the sample sequence, and the UV detector is a 2489.

FIX or WORKAROUND

Select UV1 as Fraction Trigger, but leave all the wavelength columns (Wavelength A, B, C...) blank.