M-Class UPLC stops after first injection, does not move to next one, shows "System not ready to perform operation" error - WKB88854
Article number: 88854
SYMPTOMS
- Queue stalls after first injection is complete
- µSM-FL produces "Sample Manager: System not ready to perform operation" error
- TVM produces "Running: No function or text"
- System switches to Standby mode if a shutdown method is specified
ENVIRONMENT
- MassLynx 4.2
- DP2019R2 for ACQUITY UPLC M-Class
CAUSE
Triggering is not set correctly in the Inlet Editor.
FIX or WORKAROUND
- Ensure that Event 1 IN is selected and triggering for both Pump and Detector and is set to "Trigger by Contact Closure", even though no contact closure cables are needed or present.
- Verify that upon submission of more than one sample, the queue progresses as expected.
ADDITIONAL INFORMATION
id88854, MLYNX, MLYNXV41, SUPMM, UPMASM, UPMBSM, UPMSMFL