How to confirm the matrix effect of a mass spectrometer with Fluidics (MassLynx) - WKB88063
Article number: 88063
OBJECTIVE or GOAL
Use a Fluidics pump to confirm matrix effect of the sample.
ENVIRONMENT
- MassLynx
- Xevo TQ-S
- Xevo TQ-S cronos
- Xevo TQ-S micro
- Xevo TQ-XS
PROCEDURE
- Set the standard of the target compound for Fluidics A or B.
- While the mobile phase flow, set Valve Position to Combined and confirm the intensity of the relevant MRM conditions on the Tune page.
- Set Flow State to Combined from Options (Method events) in MS Method.
- Run a blank sample using the MS Method with the Flow State changed, and confirm the ion intensity near the retention time.