No results shown in the TargetLynx browser when data are processed - WKB8794

Article number: 8794

SYMPTOMS

  • Only the information about sample appears in the TargetLynx Summary Bar (file name, File Text, Sample ID, Vial location, and so on).
  • No data or results show in the TargetLynx Summary bar (no integration, retention times, concentrations, and so on).
  • Found: Correct method is being used.
  • Found: Cannot open raw data directly from Sample list > Chromatogram option.
  • Found: There are no spaces preceding the file name in the File Name column.
  • Found: Project open in MassLynx is not the project to which the data were acquired.

ENVIRONMENT

  • MassLynx 4.1 
  • Masslynx 4.2
  • TargetLynx 4.1
  • TargetLynx 4.2
  • MassLynx

CAUSE

A mismatch exists between the project that is open in MassLynx and the sample list being processed.

FIX or WORKAROUND

  1. Go to File > Open Project and select the project to which the data was acquired.
  2. Go to File > Open and select the sample list to be processed with TargetLynx. Ensure that the sample list file is in the SampleDB folder of the project opened in Step #1.
  3. Under TargetLynx Shortcut, select Process Samples, select appropriate method, and click OK.

ADDITIONAL INFORMATION

This problem may also occur when MassLynx project folders or data file folders are moved to a different PC or a virtual PC where the file structure is different. 

Not able to find a solution? Click here to request help.