How many TargetLynx dataset (.qld) files can be imported to a TrendPlot workspace (.tpw) file? - WKB73102
Article number: 73102
ENVIRONMENT
- MassLynx
- TargetLynx
- TrendPlot
ANSWER
There is no known limit to the number of TargetLynx .qld files that can be imported into a single TrendPlot workspace (.tpw) file because this limit has not been tested.
ADDITIONAL INFORMATION
A limitation on how many .qld files can be imported into the .tpw file is more likely going to be the result of .qld file size, not the number of files. All MassLynx processes are 32-bit. There is a limitation to the amount of memory they can use. The total addressable space using 32 bit is 4 GB. However, only 2 GB is usually available for a program to allocate to memory.
TrendPlot may have difficulty if the .tpw file approaches 2 GB in size. Customers should manage their TrendPlot workspaces based on this limitation.
id73102, MLYNX, MLYNXV41, SUPMM, TARLYNX41