SYMPTOMS
- When trying to open a project in MassLynx security, receiving the following error smessage: "project not accessible"
ENVIRONMENT
- MassLynx 4.1
- MassLynx Full Security
CAUSE
The project folder has read-only permissions.
FIX or WORKAROUND
- All project folders require read, write, and execute privileges to be opened in MassLynx, Change the project folder permissions to those settings.
ADDITIONAL INFORMATION
If you want to limit who can access this project within MassLynx Security, you can set up project directories and assign only certain users to those directories.