IonLynx - "Test Name" missing in Excel report - WKB65297

Article number: 65297

SYMPTOMS

  • When reporting an excel results scheme the "Test Name" (Sample Name) is not displayed. 

IonLynx Reporting Incorrect2.png

ENVIRONMENT

  • IonLynx 

CAUSE

When selecting "Test Name" from the drop down menu it doesn't get assigned to "Calculated Conc" properly. 

IonLynx Reporting.png

FIX or WORKAROUND

When creating the excel report in the "Report Scheme Settings", make sure to do it in the exact below order. The order is absolutely critical for steps 4, 5 and 6!

1. Make sure the "Report Format" section is setup according to the below picture (all boxes checked, Delimiter set to <tab>)

2. Select "File Name" on the left hand side menu of the "Sample Report Header" Section

3. Add "File Name" to the right hand side by clicking the plus sign

4. Select "Calculated Conc" on the left hand side menu of the "Sample Report Header" Section

5. Add "Calculated Conc" to the right hand side by clicking the plus sign

6. From the "Header" drop down menu select "Test Name"

7. In the "Report Path" section setup the output file name and location

NeoLynx Excel Reporting.png

ADDITIONAL INFORMATION

See picture blow for correct excel output: 

IonLynx Reporting Correct.png