Error "Warning: NIST MS Database directory invalid" - WKB590
Article number: 590
SYMPTOMS
- Error appears when attempting to search NIST libraries:
ENVIRONMENT
- MassLynx 4.1
- NIST 2011
CAUSE
Library Search List points to the directory of the NIST 2011 libraries (C:\NIST2011), not the actual library files.
FIX or WORKAROUND
- In the main MassLynx window, enable Shortcuts.
- Go to the Tools menu.
- Select Search Library.
- In the Library window that opens, select its File menu.
- Choose Search List.
- In the Library Search List section, highlight C:\NIST2011 and click the Remove button.
- Click the Add button.
- Use the Browse For Folder window to navigate to C:NIST2011\MSSEARCH.
- There are four libraries here: mainlib, nist_msms, nist_ri, and replib. Highlight the folder with the pertinent library name and click OK.
- Repeat steps 7 through 9 if more than one library is involved.
- Click OK to exit.
ADDITIONAL INFORMATION
id590, MLYNX, MLYNXV41, SUPMM