Cannot import MS method into QuanPedia - WKB5896

Article number: 5896

SYMPTOMS

  • Cannot import MS method into QuanPedia.
  • Error message: Failed to import.

ENVIRONMENT

  • MassLynx 4.1 SCN 945 with SCN 960
  • QuanPedia

CAUSE

MS method fails to import if it contains an MRM transition that is not present in the associated TargetlLnx method.

FIX or WORKAROUND

  1. Resolution for importing MS methods into QuanPedia found in the following two attachments.