"Number of samples must be a multiple of four" message when queuing up a single sample - WKB58922

Article number: 58922

SYMPTOMS

  • "Number of samples must be a multiple of four" message when queuing up a single sample
  • Was previously able to queue up individual samples

ENVIRONMENT

  • MassLynx 4.1

CAUSE

Multi-Probe Capability is enabled.

FIX or WORKAROUND

  1. Under the Instrument Shortcut, go to Options > Multi-probe tab.
  2. Clear the selection for Multi-Probe Capability, and click OK.

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.