How to resolve MS system IP conflicts with customer LAN - WKB55039
OBJECTIVE or GOAL
Resolve IP conflicts between customer LAN and system LAN.
ENVIRONMENT
- MassLynx 4.1
- MassLynx 4.2
- All mass spectrometers
- Empower 3
PROCEDURE
-
Reconfigure the Waters DHCP Wizard to use an alternative IP address (e.g., 172.16.0.1). (This will release the current entries and automatically change the address of the network card.)
-
Reboot the MassLynx PC and reboot the instrument stack.
-
Re-establish communication between the MassLynx PC and the instrument stack.
-
If communications are not re-established, make sure that the PC recognizes the new EPC IP address:
-
In the search box, type "ping epc".
-
If the address pinged for EPC is not the correct IP address, edit the hosts file to the new IP address.
-
Edit the IP hosts file, C: drive > Windows > System32 > drivers > Etc > hosts.
-
Open the file "host " with Note Pad, change the IP address in this file to, for example, 172.16.0.x, where x is the specific IP address of the mass spectrometer.